The Horiba OCMA-550 analyzer simplifies measurement of oil in water or soil for a wide variety of applications. Use this precision instrument for surveying water quality and hazardous waste sites, industrial wastewater monitoring, post-cleaning discharge analysis, and measuring adherent oil residuals on textiles and metal parts.
Use S-316 solvent to quantitatively extract sample. The oil / solvent solution is then simply transferred to the OCMA-550 analyzer for single-button measurement. Analysis is based on infrared (IR) spectral data, with peak integration (at approximately 3.4 µm) to determine oil concentration. S-316 solvent is non-absorbant at this IR wavelength, so only oil is measured. Accurately measures oil down to 0.1 ppm. Auto-calibrates after initial standard reading.
Use built-in time stamping to document results. Read results on LCD or transfer data using parallel printer port or to computer via RS-232. S-316 solvent can be recycled using Solvent Reclaimer SR-300 (82900-20), sold separately below.
What’s included: Quartz cell with cap, power cord, B-heavy oil (standard), 2 amp fuse, 10 mL syringe and 25 µL syringe.