技术参数:
Water-to-air cooling unit
Cooling capacity up to 2 kW (0.5, 1, 1.5, 2 kW options available)
Laser loop coolant flow up to 7 l/min
Pressure up to 3.5 bar
Coolant temperature indication
Water filtration and deionisation
19 x 7U case
25-45°C temperature range
COOLING UNIT PS1245CO is designed for flashlamp-pumped lasers and provides effective heat removal and high stability of laser rod temperature. This enables long term operation of your laser at maximum efficiency. Unit is assembled in 19" chase and may be used completing the powering group of laser systems. Also it can be mounted into the 19 rack.
Operation of cooling unit is based on water-to-air heat exchange. Deionised or distilled water of 1-20 µS/cm conductivity is used in laser loop of cooling unit.
Temperature stability of coolant is maintained by electronic circuit. Stabilization temperature is set by 10-turn potentiometer within 25-45°C (another range is available according to customers requests).
Indication of coolant and set temperature, coolant level, overheat and pressure absence in laser loop is provided on front panel of cooling unit.
Blocking and protection circuitry in case of overheating, pressure absence or coolant level is out of low limits sends interrupt signal to interlock connector to which protection circuitry of laser powering might be connected. Coolant pump will be stopped in case of pressure absence or coolant level is out of limits.
Deionisation and filtration of coolant eliminates a build up on the flashlamp, reduces servicing need and allows laser operation at maximum efficiency in day-to-day use.