自动运动粘度试验仪
AKV-202
Model AKV-202 has been designed for automatic determination of KINEMATIC VISCOSITY. Once a specimen is set the series of processes including timing, cleaning/drying of viscometer tube, and reporting are fully automated. AKV-202 is an economical bench-top system, which saves not only the technician’s time but also valuable laboratory floor space and budget.
AKV-202自动运动粘度测试仪是用于自动测定石油产品运动粘度的试验装置。只需要简单设定就可以自动完成从样品进样,样品测试,打印试验结果,到清洗并干燥粘度计的一系列工作。AKV-202是一个设计精巧的台式试验装置,占用空间小,经济适用。试验过程完全自动,既避免了人为因素的干扰,也节约了试验操作人员大量的时间。
WIDE RANGE OF SAMPLES: Wide range of Kinematic Viscosity(KV) can be covered with the modified Lanz-Zeitfuchs viscometers. The reverse flow viscometer covers both transparent and opaque samples and also eliminates drainage error.
粘度测定范围广:采用了改进型兰兹-斯特福斯粘度计,可以测定各种透明和不透明试样。
AUTOMATIC MEASUREMNT: Timing, sample draining, viscometer cleaning/drying, calculating/reporting are automated, saving the time of the experienced hands and improving repeatability.
自动测试:计时,自动进样和清洗粘度计,计算结果并打印报告的过程全自动进行,可节省大量时间,无须人员的手工操作和熟练度,试验结果可靠和有重复性。
COMPACT DESIGN: The two-viscometers/one-bath system is housed in a compact housing that is small enough to install on a lab bench.
紧凑的设计:台式试验仪器,封闭式恒温浴槽,放置于实验台上,可同时进行2支(同组或分组的)运动粘度测试
EASY VISCOMETER REPLACEMENT: Viscometer is mounted to a viscometer holder and the viscometer/ holder assembly can be removed from the bath at ease, which assures easy viscometer replacement.
粘度计用独立的支架安装,可以简单的取出和更换
OVERFLOW: The viscometer has a provision of overflow, which eliminates errors from inconsistency in samp领 volume as well as thermal expansion of the sample.
改进型兰兹-斯特福斯粘度计可有效的减少热胀冷缩对试验的影响
VERSATILE CLEANING/DRYING SYSTEM: Two tanks are provided for cleaning and drying the viscometers, which allows quick and efficient cleaning/drying by selecting the best chemicals for the sample.
独有的清洗系统:可选用2种不同的化学清洗液对粘度计进行清洗,清洗效果好,无残留,清洗完成后还可自动吹风干燥粘度计。
TANAKA Kinematic Viscosity Measuring System, model AKV-202
TYPE
Bench top automatic KV measuring system with 2 viscometers in 1 bath
自动运动粘度测试,封闭式恒温浴槽,2支粘度计
RELATED STANDARDS: (适用标准)
ISO 3104, ASTM D445, IP71, GB/T 265 etc.
VISCOMETERS: (粘度计类型)
Modified Lanz-Zeitfuchs viscometers 改进型兰兹-斯特福斯粘度计
MEASURING RANGE: (测试范围)
10,000 mm2/s for automatic operation. (Range depends on viscometer.)
10,000 mm2/s 根据下表选择不同的粘度计
MENISCUS DETECTION: (检测方式)
By photo-electric devices with fiber optics
光纤传导 + 光电检测传感器
TIMING: (时间计数)
0.00 - 999.99 sec (0.00 - 999.99 秒,精确到2位小数)
BATH: (恒温浴槽)
Bath: Stainless Steel bath with drain, 18 liters Heater: Sheathed heaters, 0.59kW x2 Lighting: Fluorescent light, 6W
全不锈钢恒温浴槽,约 18升,内置加热装置(0.59kW x2),后置荧光照明灯(6W)
Medium:Water(distilled or ion exchanged)for < 75℃,
75℃以下时可用蒸馏水或去离子水作为恒温介质
Colorless/transparent silicon oil (with PMCC flash point of 190 ℃ at minimum and KV of 10-20mm2/s) for >= 75 ℃.
75℃以上时用无色透明的硅油作为恒温介质(要求PMCC闭口闪点大于190 ℃,运动粘度为10-20mm2/s)
BATH TEMPERATURE CONTROLLER: (浴槽恒温控制)
Temp: Preset at 6 points. Preset point can be set at any temp between 20 and 100. ℃ Precision: +/-0.01 ℃ Controller: PID control by a micro-processor
20℃100℃,预设6点,PID控温精度±0.01℃
CLEANING/DRYING VISCOMETERS: (粘度计的清洗和干燥)
An automatic cycle consisting of draining sample, injecting chemicals, and drying with air. By vacuum.
在每次测试后,按程序注入清洗溶剂,进行多次负压牵引清洗,排出废液后进行吹风干燥粘度计
Chemicals: BTX for cleaning, Ethanol (Bath Temp <80oC)for drying. (Ether or Ketone can not be used)
清洗液:在恒温浴槽温度小于80℃的情况下,可选用BTX (对二甲苯)和乙醇作为清洗溶剂(不可使用燃点较低的醚类或酮类化学品)
Fume: Fume to be exhausted through trap and hose to outside
废气可通过软管引出室外排放或用收集瓶收集。
DISPLAY: (显示输出)
Test parameters, status, trouble, etc. displayed on fluorescence display module
荧光数显:测试参数,测试结果,出错警示,等。
PRINTER: (打印机)
Thermal printer, 58mm of paper width 热敏打印机,纸宽 58mm
COMPUTER OUTPUT: (计算机输出)
RS-232C, 1 channel RS-232C标准接口 1
SAFETY: (安全保护)
Alarms for gas, bath medium level, & bath overheat 气体保护,液位保护,温度保护
POWER REQUIREMENTS: (电源需求)
100, 120, 220 or 240VAC, 2KW
DIMENSION & WEIGHT: (体积/重量)
530mmW x 560mmD x 930mmH, 75kg
ORDERING INFORMATION: (定购要求)
STANDARD ACCESSORIES 标准附件
1.Viscometers(*) 粘度计
2
2.Viscometer holders粘度计支架
2
3.Beaker样品杯
5
4.Clips for spherical joints 球面夹箍
4
5.Printer roll papers打印纸
1
6.Duct hose 连接软管
5m
7.Cleaning solvent tanks 清洗液瓶
2
8.Overflow tank 废液瓶
1
8.Drain jars 排液桶
2
9.Exhaust gas trap 废气冷凝收集瓶
1
*Select 2 viscometers from the table below: (按下表范围任选2支粘度计)
Tanaka Viscometer No.
KV range mm2/s
TLZ 1
0.75 3.75
TLZ 2
2 -10
TLZ 3
6 -30
TLZ 4
20 -100
TLZ 5
60 -300
TLZ 6
200 -1,000
TLZ 7
600 -3,000
TLZ 8
2,000 -10,000
OPTIONAL ACCESSORIES: (推荐附件)
1.Thermometers for bath: for 30, 40, 50, 80 &100 ℃
检验用温度计,30, 40, 50, 80 &100℃
2.Chiller: Used when differential between room temp. and bath temp. is less than 15 ℃. Thomas TRL-108H (Made in Japan) or Julabo F200 (Made in Germany) is recommended.
制冷源:试验要求和室温环境相差15℃以上时,需要加配外接制冷源。
推荐:TRL-108H (日本田中) Julabo F200(德国)
3.Silicon oil for bath (17 kg) when bath temperature is set 80 ℃ or higher.
无色高温硅油,用于80 ℃以上的运动粘度试验,17 kg装