TheEnduraAZ20isthelatestinalonglineofhigh-quality,combustiongasanalyzersfromABB.
Thesensor,basedonazirconiumoxidecell,ismountedatthetipoftheprobethatisinsertedintheflueduct.Theresultingdirect,insitumeasurementprovidesaccurateandrapidoxygenreadingforcombustioncontrol/optimizationandemissionsmonitoringpurposes.
EnduraAZ20hasbeendesignedforextendedperiodsofmaintenance-freeoperation.Themodulardesignwithreducedcomponentcountimprovesthereliabilityandrobustnessofthesystemandsimplifiesbreakdownrepairifitoccurs.
Kitscontainingallthepartsneededtocompleteon-siterepairsareavailablefromABB,ensuringthatservicepersonnelcaneffectrepairsquicklyandefficientlyatminimumcost.TheEnduraAZ20probehasretainedaneasy-accesscellarrangement,similartotheprovenZFG2probe.Thisensurescellreplacementcanbecarried
outon-siteusingreadilyavailable,basichandtools,evenafterlongperiodsofhightemperatureoperation.
TheEnduraAZ20oxygenprobemeasuresoxygenconcentrationinfluegasusinganinsitu'wetanalysis'method.The'wetanalysis'methodavoidsmeasurementerror(typically20%ofreadinghigherthantheactualvalue)thatisintroducedbyasamp领systemusingthe'dryanalysis'method.
ForcustomerswithanexistinginstalledZFG2probewithaZMTorZDTtransmitter,a55VheaterversionoftheEnduraAZ20probe(theEnduraAZ20/ZFG2replacementprobe)isavailable.
Systemequipmentcomprisesa(flue-mounted)EnduraAZ20probecontrolledbyanintegralorremotetransmitter.Duringoperation,azirconiacellwithintheprobeisheldataconstanttemperatureof700°C(1292°F)byaprobeheaterandcontrol
thermocoupleassembly.Iftheheatercontrolcircuitryorsoftwarefails(unsafe),theprobeheaterpowersupplyisswitchedoff,ensuringthesystemfails'safe'andprotectstheheaterfromanover-tempfailure.
AnoutputgeneratedatthezirconiacellisprocessedinthetransmittergivingalocallydisplayedO2readinganda4to20mAretransmissionsignaloveranyrangebetween0%and1**%O2.
Optionalautomaticcalibration(AutoCal*)enablesautomatic,semi-automaticormanualcalibrationtobeperformedusingaprobe-mountedgascontrolmanifoldmountedwithintheprobehead.Calibrationsequencingissoftware-controlledfromthetransmitter.
Optionalrestrictors(wherefitted*)intheprobeheadcontrolthetestgasandreferenceairflowwithouttheneedforflowmeters.Theproberequiresonlypresettestgasandreferenceairpressuresof1bar(15psi)tomaintainaconstantflowof2.2l/min(4.662scfh).Thisflowisnotaffectedbychangesinthemeasuredgaspressureof±0.35bar(5psi).
Introduction
TheEnduraAZ25combustionoxygenanalyzeristhelatestinalonglineofhigh-quality,combustiongasanalyzersfromABB.TheAZ25isdesignedtomeasureoxygenconcentrationinhightemperaturefurnaceatmospheres.
Thesensor,basedonazirconiumoxidecell,ismountedatthetipoftheprobethatisinserteddirectlyintothefurnace.Theresultingdirect,in-situmeasurementprovidesanaccurateandrapidoxygenreading,forcombustioncontrol/optimizationandprocessqualitycontrolpurposes.
Thein-situtechniqueprovidesatrueoxygenmeasurementinsidetheprocess.Iteliminates'equilibriumshift'thatiscommontosystemsinwhichgassamplesareextractedandcooledbeforemeasurement.
Referenceair(500to1000ml/min[1to2ft3/hr])isrequiredforaccurateoperationandcanbesuppliedfromanABBmains-poweredpumpunitoraflowregulatorunit.
Uniqueintegratedautomaticcalibration
TheEnduraAZ25automaticcalibrationsystemeliminatestheneedfortheexpensiveancillaryequipmentrequiredforautomaticcalibrationontraditionalfluegasoxygenanalyzersystems.ABB’sfullyintegrated,automaticcalibrationfeature
controlsthetestgassequenceanddetectstestgasavailability,eliminatingincorrectcalibrationsduetolossoftestgas.Integratedautomaticcalibrationsecurestheoxygenmeasurement,providingcompleteconfidenceintheanalyzer's
performance.
Long-termoperationwithouttechnicianinterventionminimizesoperatingcosts,totalcost-of-ownershipandoptimizessystemaccuracy.Asignificantproportionofservicevisitstoananalyzerresultin'nofaultfound'.EnduraAZ25'shigh-quality
manufacture,advanceddiagnosticsandautomaticcalibrationistargetedatreducingthiswastedeffortandcost.
Dependable
measurementofO2inawidevarietyofprocesses
virtuallyinstantaneousresponse
resistancetosulphurousreducingatmospheres
Versatile
600to1250mm(2to4ft)lengths
choiceofprotectivesheathmaterials
integralthermocoupleforautomatictemperature
compensation
hightemperatureoperationupto1400°C(2552°F)
Lowcost-of-ownership
insitumeasurement
nosamp领systemrequired
automaticcalibrationfunctionalityasstandard
optionalautomaticcalibrationhardware
Applications
Furnaceatmospheremonitoring:
Annea领furnace
Sinteringofmetals
Ferritesintering
Whiteheartmalleablizing
Combustionmeasurementandcontrolin:
Soakingpits
Processheaters
Fluidizedbedboilers
Hightemperatureincinerators
Allothercombustionprocesseswherethefluegastemperatureisinexcessof600°C(1112°F)
Aluminousporcelain
Entrylevelsheathmaterialsuitableforapplicationswithtemperaturesupto1250°C(2282°F)limitedresistanttocorrosivechemicalsandmechanicalandthermalstress.
Recrystallizedalumina
Suitableforapplicationswithtemperaturesupto1400°C(2552°F)resistanttocorrosivechemicalsandmoreseveremechanicalandthermalstressarepresent.
Incoloy800
Suitableforapplicationswithtemperaturesupto850°C
(1562°F)whenmountedhorizontallyandupto1000°C
(1832°F)whenmountedvertically.
Offeringlimitedresistancetocorrosivechemicalsandresistantmechanicalandthermalstress.
爱泽工业(IZE_INDUSTRIES)销售型号示例:
ABB氧化锆分析仪AZ25/.1.1.2.1.1.5.3.1.2.0.0.0.0/STD
ABB氧化锆分析仪AZ25/.1.1.2.1.1.5.1.1.1.1.0.0.0.0
ABB氧化锆分析仪AZ25/.1.1.2.1.1.5.1.1.1.1.0.0.0.0.E./STD
ABB氧化锆分析仪AZ25/.1.1.2.1.1.5.3.1.2.0.0.0.0.E.Z.J/STD
ABB氧化锆分析仪AZ20/112112225111100C/STD
ABB氧化锆分析仪AZ20/1.2.2.2.1.2.2.3.3.7.1.1.2.E
ABB氧化锆分析仪AZ20/222122337112E
ABB氧化锆分析仪AZ20/2.2.2.1.2.2.3.3.7.1.1.21.0.0.E/STD
ABB氧化锆分析仪AZ20/222122337112E/STD
如果您有任何关于产品的相关技术咨询或者需要产品的相关技术资料与文档,请随时联系爱泽工业(86-21-31006702),或者访问网站www.ize-industries。。com,我们将尽全力解决您的问题。
Ifyouhaveanyquestion,pleasefeelfreetodial0086-21-31006702orbrowseourwebsite:www.ize-industries。。com.Thankyou.
仪器网-专业分析仪器服务平台,实验室仪器设备交易网,仪器行业专业网络宣传媒体。
相关热词:
等离子清洗机,反应釜,旋转蒸发仪,高精度温湿度计,露点仪,高效液相色谱仪价格,霉菌试验箱,跌落试验台,离子色谱仪价格,噪声计,高压灭菌器,集菌仪,接地电阻测试仪型号,柱温箱,旋涡混合仪,电热套,场强仪万能材料试验机价格,洗瓶机,匀浆机,耐候试验箱,熔融指数仪,透射电子显微镜。